MESA/Dose Info Consumer
From IHEWiki
Revision as of 18:08, 16 October 2008; Felhofer (Talk | contribs)
(diff) ←Older revision | Current revision | Newer revision→ (diff)
(diff) ←Older revision | Current revision | Newer revision→ (diff)
MESA/Dose Information Consumer
Contents |
[edit]
Radiation Exposure Monitoring Integration Profile Tests
This section describes tests that are specific to the IHE Radiation Exposure Monitoring Integration (REM) profile.
[edit]
Dose Information Consumer Test 4400: REM Vendor Interoperability
Dose Information Reporters in the REM profile are required to support storage of DICOM Dose SR objects.
This test gives you access to sample DICOM SRs created by REM Acquisition Modalities prior to the Connectathon. Modalities are asked to submit samples into kudu 3 weeks prior to the MESA test deadline (or earlier). The goal is to avoid surprises on the Connectathon floor.
- You will find these samples in kudu under MESA Tests -> Pre-Connectathon Images.... Select your system name and look under Objects to render to find sample Dose SRs for test 4400.
- Your task is to test with these samples prior to the Connectathon. Do not complete this test before those samples are available.
- Create a short txt file indicating you have accessed these samples. List those that you have successfully tested with your Dose Information Consumer. If you encounter problems, contact the Project Manager who will help resolve them with Modality vendors. Upload that txt file into the Kudu system under the MESA test log results for this test.